Correlation Studies
Most laboratories doing microscopy have "self-calibrated" to
some internal standard. Such self-calibration is precisely
the target of CLIA and other regulatory mechanisms, but
is a regrettable reality in sperm analysis.
Until the advent of the SQA, it has not been possible to use an objective standard,
so each population of laboratorians has converged upon its own set of judgment
rules.
Because of that self-calibration, it is almost certain that when the SQA is introduced
to a lab there will be significant differences between its output and that of
microscopic evaluations. Fortunately, this is easy to resolve once it is accepted
that the lab is offset by some factor.
Whether or not the results match, the issue is whether
they are coherent and monotonic. That is, when one
goes up so should the other, and vice versa. The
animated graph below is an example of "good" correlation.
Note that on that initially the tests are offset
one from the other, and in no case did both tests
achieve the same result.
 |
They
"correlate" because they follow one another
closely. By applying a simple offset correction, they
can be made to look almost identical. |
 |
On the left is an example
of "poor" correlation. Some of the test pairs
achieved exactly the same results, but others were
so far off that a statistical analysis shows very poor
performance, and there is no offset correction that
would improve the fit between the curves. |
The following figures reflect the results of studies by
a reputable laboratory, after a statistical analysis applied
the required offset. We have been very critical in presenting
this information: there are better results achieved by
many SQA users, but these correlations are typical.
You will note that in this particular set there was a problem in evaluating samples
4 and 5. Later analysis indicated that the error was in the microscopy rather
than the SQA.
|